- GB10 Gebruiksaanwijzing
The ADL5 is widely used for measuring thickness and corrosion of pressure vessels, chemical equipment, boilers, oil storage tanks, etc. In industries of petroleum, shipbuilding, power station, and machine manufacturing.
Operating principle:
F type - (magnetic induction), measure the thickness of non-magnetic coating layers such as (aluminum, chrome, copper, ename1, rubber, paint) on magnetic substrate (steel, iron, alloy and magnetic steel).
NF type - (eddy current), measure the thickness of non-conductive coating layers (ename1, rubber, paint, vanish, plastic anodic oxide layer) covered on non-ferrous substrate (aluminum, brass, zinc, tin and non-magnetic stainless steel).
| Measuring range: | 0-3500um |
| Resolution: | 0.1/1 |
| Accuracy: | +/-1-3%n or +/-2.5um |
| Min. measuring area: | 6mm |
| Min. sample thickness: | 0.3mm |
| Battery indicator: | Low battery indicator |
| Metric/imperial: | Convertible |
| Power supply: | 4pcs. x1.5V AAA(UM-4)battery; Auto power off |
| Operating conditions: | 0-+45C(32F-104F), ≤90%RH |
| Dimensions: | 126 x 65 x 27mm |
| Weight: | 81grams(not including battery) |
| Goedkeuring: | CE, RoHS en WEEE |